Targeted dual-energy single-source CT for characterisation of urinary calculi: experimental and clinical experience.
M. Eiber, K. Holzapfel, M. Frimberger, M. Straub, H. Schneider, EJ. Rummeny, M. Dobritz, A. Huber
Eur Radiol 2012 Jan;22(1):251-8.
PubMed: 21847542
Abstract
To assess the accuracy of targeted dual-energy single-source multi-detector CT (MDCT) for characterisation of urinary calculi. For proof of principle, 71 ex-vivo calculi underwent single-source 256-slice MDCT. Low-dose CT was performed in 154 patients with suspected urinary calculi. In 104 patients with urinary calculi targeted dual-energy imaging within one breath-hold was added. 46 patients with sufficient material for infrared-spectroscopy were analysed. Potential anatomical misregistrations between 80- and 140-kV(p)-images and HU-values were measured. DEIs (dual-energy-indices) were compared with the standard of reference. Effective doses were calculated. In 26 of 46 patients no misregistration was present. Mean deviations were 2.7 mm in the z-axis (16 patients) and 4.3 mm in the axial plane (10 patients). The DEIs were 0.018 ± 0.016 for uric acid (UA), 0.035 ± 0.015 for mixed UA and 0.102 ± 0.015 for calcified stones in-vitro and 0.017 ± 0.002 for UA, 0.050 ± 0.019 for mixed UA and 0.122 ± 0.024 for calcified calculi in-vivo. Significant differences were noted among calcium, mixed UA and UA stones (p